A liver scan can report how stiff the liver is, in kilopascals. The machine does not press on the organ and feel it. It gives the tissue a small push, sends a slow sideways wave through it, and times that wave with ordinary ultrasound pulses. The ultrasound itself travels at about 1,540 m/s. The sideways wave crawls along at a metre or two per second. That slow speed is the whole measurement, and the physics behind it is the wave-speed formula from Class 11.
This post is about the physics only. It gives no medical advice, and the numbers in it do not judge anyone's health. Only a doctor can interpret a real scan.
Why can't ordinary ultrasound see how stiff tissue is?
Soft tissue is mostly water. Water is very hard to squeeze into a smaller volume, but very easy to shear (slide one layer over the next). The two kinds of wave in tissue feel these two different stiffnesses.
Longitudinal (sound) waves squeeze and stretch the tissue along the direction of travel. Their speed is c_L = √(K/ρ), where K is the bulk modulus and ρ the density.
Shear (transverse) waves move the tissue sideways. Their speed is c_s = √(G/ρ), where G is the shear modulus.
Solids can carry the second kind and gases cannot, as explained in why transverse waves cannot travel through gases. Soft tissue is a very soft solid, so it carries shear waves, but slowly.
With c_L = 1,540 m/s and ρ = 1,000 kg/m³, K = ρc_L² = 1,000 × 1,540² ≈ 2.37 × 10⁹ Pa, or 2.37 GPa. Every soft tissue has nearly the same sound speed, so K hardly changes from one tissue to the next. An ordinary scan, which uses sound speed and echoes (see how an ultrasound scan makes a picture from echoes), cannot tell a soft organ from a stiffer one this way.
The shear modulus is the opposite. The WFUMB guidelines on the basic principles of ultrasound elastography (Shiina and co-authors, Ultrasound in Medicine and Biology, 2015) give shear wave speeds of roughly 1 to 10 m/s in soft tissue, against 1,540 m/s for sound, and shear moduli of roughly 1 to 100 kPa. G varies a lot between tissues, so it gives good contrast.
How do you get Young's modulus from the shear wave speed?
The machine measures c_s. Squaring the wave-speed formula gives G = ρc_s². Most machines report Young's modulus E instead, using the elastic-constant relation
E = 2G(1 + ν)
where ν is Poisson's ratio. For a material that keeps its volume when it is squeezed, ν = 0.5, and E = 3G. The same WFUMB guideline writes the working formula as E = 3G = 3ρc_s², and says the density is usually taken as 1 g/cm³ (1,000 kg/m³).
Why is ν so close to 0.5? Poisson's ratio can be written in terms of K and G as ν = (3K − 2G) / (2(3K + G)). Take K = 2.37 GPa and G = 2.25 kPa. Then K/G is about a million, and ν comes out as 0.4999995. So E = 2G(1 + ν) is 3G to better than one part in a million. If you want the meaning of ν for a stretched wire first, see Poisson's ratio and the volume change of a stretched wire.
How does the machine time a wave that slow?
There are two ways to make the push. In transient elastography, a small piston on the probe taps the skin once. In acoustic radiation force methods, a long, focused ultrasound pulse pushes the tissue from inside. Either way, a shear wave spreads out from the push.
The probe then fires short ultrasound pulses very often and compares each echo with the one before. Tissue moved by the shear wave shifts the echo slightly, so the machine sees where the wave front is at each moment.
Check that the timing is easy. To image 4 cm deep, a pulse goes 4 cm down and 4 cm back, 0.08 m in total. At 1,540 m/s that takes 0.08 / 1,540 = 52 µs, so the probe could in principle fire about 19,000 pulses a second. Even at an illustrative 5,000 frames a second, a shear wave at 1.5 m/s moves only 1.5 × 0.0002 = 0.0003 m, or 0.3 mm, between frames. Crossing 40 mm takes it 0.040 / 1.5 = 26.7 ms.
Worked example: from wave arrival times to kilopascals
Suppose the shear wave front reaches a depth of 25 mm at 8.0 ms after the push and a depth of 55 mm at 28.0 ms. These are illustrative numbers. Take ρ = 1,000 kg/m³, as the machines do.
Distance between the two points. Δx = 55 − 25 = 30 mm = 0.030 m.
Time between them. Δt = 28.0 − 8.0 = 20.0 ms = 0.020 s.
Shear wave speed. c_s = Δx / Δt = 0.030 / 0.020 = 1.5 m/s.
Shear modulus. G = ρc_s² = 1,000 × 1.5² = 2,250 Pa = 2.25 kPa.
Young's modulus. E = 3G = 6,750 Pa, or about 6.8 kPa.
Check the units. (kg/m³) × (m/s)² = kg/(m s²) = N/m² = Pa. Good.
Check the size. 2.25 kPa sits inside the 1 to 100 kPa range in the guideline.
Because E grows as c², doubling the speed makes E four times as large:
1.0 m/s gives E = 3.0 kPa.
1.5 m/s gives E = 6.8 kPa.
2.0 m/s gives E = 12 kPa.
3.0 m/s gives E = 27 kPa.
These are physics examples, not reference values. If the real tissue density were 1,060 kg/m³ rather than 1,000 kg/m³, the 1.5 m/s case would give 7.2 kPa instead of 6.8 kPa, a 6% change.
What makes the real measurement harder than the formula?
E = 3ρc_s² assumes the tissue is elastic, uniform, the same in every direction and nearly incompressible. Real tissue is also viscous, so the measured speed depends a little on the frequency of the shear wave. The guideline notes that shear waves die out quickly at megahertz frequencies and only travel usefully in the body below about 1 kHz. At an illustrative 50 Hz, a 1.5 m/s wave has a wavelength of λ = c/f = 1.5 / 50 = 0.030 m, or 30 mm, so the machine measures an average over a few centimetres, not a single point.
Inferring an artery's stiffness from the pulse wave velocity uses a wave's speed in the same way, for a wave along a tube.
If you want to see how transverse and longitudinal waves move through a medium, try the Simulation Lab, or revise elastic moduli and wave speed in the Concept Library.
Quick checklist
Sound in tissue: c_L = √(K/ρ) ≈ 1,540 m/s, K ≈ 2.4 GPa, nearly the same in every soft tissue.
Shear waves in tissue: c_s = √(G/ρ), roughly 1 to 10 m/s, G roughly 1 to 100 kPa.
Nearly incompressible, so ν ≈ 0.5 and E = 2G(1 + ν) = 3G = 3ρc_s².
E scales as c²: twice the speed means four times the stiffness.
Units check: (kg/m³)(m/s)² = Pa.
This is physics only. A stiffness number is read by a doctor, never by the formula alone.