On this page
In an intensive care unit, doctors sometimes need to know how much blood the heart pumps each minute. One way uses a syringe of cold salt water and a tiny thermometer. A clinical team injects about 10 mL of saline into a vein near the heart, and a sensor further downstream records how much the blood cools, and for how long. The method is called thermodilution, and the physics behind it is the method of mixtures from your calorimetry chapter. This post explains the physics only; these measurements are planned, done and interpreted by trained clinical teams.
What is thermodilution and how does it work?
A long, thin catheter is passed through a vein into the right side of the heart, with its tip in the pulmonary artery, the vessel that carries blood from the heart to the lungs. It has two parts that matter here:
An injection port that opens in the right atrium, where the cold saline goes in.
A thermistor near the tip, in the pulmonary artery, that records blood temperature many times a second.
The cold saline mixes with blood as the right ventricle pumps it onward. When the cooled blood flows past the thermistor, the temperature dips, then recovers as warm blood replaces it. The size and length of that dip tell you the flow.
Thermodilution was introduced by Fegler in 1954 (Nishikawa and Dohi, Canadian Journal of Anaesthesia, 1993), and it became a bedside method in the early 1970s after Swan and Ganz developed the pulmonary artery catheter (Argueta and Paniagua, Cardiology in Review, 2019).
Why does a smaller temperature dip mean more blood flow?
Think of the cold saline as a fixed amount of "cold" to be shared out. Warming up to body temperature, it takes a fixed amount of heat from the blood, set by the saline alone: its volume, specific heat and starting temperature.
Now ask who supplies that heat. Every millilitre of blood that flows past during the dip gives up a little heat and cools by ΔT.
High flow: a lot of blood shares the same heat, so each millilitre cools only slightly, and the cold slug is swept past quickly. The dip is small and short.
Low flow: less blood shares the same heat, so each millilitre cools more, and the slug lingers. The dip is deep and long.
So the flow is inversely proportional to the area under the temperature-dip curve. This is the same balance as dropping a cold ball into a beaker in the method of mixtures, except that the "water" keeps flowing.
How do you derive the thermodilution equation?
Write the method of mixtures as heat gained by saline = heat lost by blood.
Heat gained by saline: V_i ρ_i c_i (T_b − T_i), where V_i is the injected volume, ρ_i c_i is the saline's heat capacity per millilitre, T_b is blood temperature and T_i is the injectate temperature.
Heat lost by blood: in a short time dt, a volume Q dt flows past cooled by ΔT, so it gives up ρ_b c_b Q ΔT dt. Add this over the whole dip: ρ_b c_b Q ∫ΔT dt.
Set them equal and solve for Q:
Q = V_i (T_b − T_i) K / ∫ΔT dt, with K = ρ_i c_i / ρ_b c_b
This is the Stewart–Hamilton equation written for heat. With approximate values, saline carries about 4.1 J per mL per kelvin and blood about 3.8 J per mL per kelvin, so K ≈ 1.08. Real monitors use a computation constant supplied by the catheter maker, which also allows for heat the saline picks up inside the catheter before it reaches the blood.
Worked example: 10 mL of saline at 0 °C
Take V_i = 10 mL, T_i = 0 °C, T_b = 37 °C, ρ_i c_i ≈ 4.1 J mL⁻¹ K⁻¹ and ρ_b c_b ≈ 3.8 J mL⁻¹ K⁻¹. These are approximate values for illustration.
Heat the saline takes up. 10 × 4.1 × 37 ≈ 1,517 J, about 1.5 kJ.
Convert a cardiac output of 5.0 L/min to mL/s. 5,000 / 60 = 83.3 mL/s.
Area under the dip. ∫ΔT dt = 1,517 / (3.8 × 83.3) ≈ 4.79 K s.
Check by reversing. If the monitor measures an area of 4.79 K s, then Q = 10 × 37 × 1.08 / 4.79 ≈ 83.3 mL/s, which is 5.0 L/min.
Other flows. At 2.5 L/min the area doubles to about 9.58 K s. At 8.0 L/min it falls to about 2.99 K s.
Size of the dip. If the 5 L/min dip is spread over roughly 5 to 6 seconds, a peak of under 1 °C is enough to give 4.79 K s. In the illustrative curve in the diagram above, the peak is about 0.9 °C.
Check the units: mL × K divided by K s gives mL/s, a flow rate.
Why use cold saline rather than room-temperature saline?
The signal is the temperature difference. At 22 °C, the same 10 mL carries 15 K of "cold" instead of 37 K, so at 5 L/min the area falls to about 1.94 K s, about 40% of the iced value. Normal blood temperature wanders slightly with breathing, so a smaller dip is harder to separate from that background.
Both are used. One review recommends repeated measurements with 10 mL at room temperature for adults, taken at evenly spaced points in the breathing cycle, and it lists conditions such as very low output and leaking tricuspid or pulmonary valves that can make the result unreliable (Nishikawa and Dohi, 1993). The choice of injectate and the reading of the curve belong to the clinical team.
What assumptions does the method make?
Complete mixing. All the saline mixes with the blood passing the thermistor.
A single pass. Cooled blood returns to the thermistor after one trip round the body. Monitors handle this by fitting the falling part of the curve and extending it, so the recirculated part is left out.
No heat lost elsewhere. Heat taken up from the catheter walls and nearby tissue is folded into the device constant.
Steady flow. Output stays constant during the measurement.
Two other ways to find cardiac output are worth knowing. You can estimate it from stroke volume and heart rate, as in how much power the human heart produces, and the biology post on cardiac output and heart rate calculations works that route in detail. Dye dilution, which how a pulse oximeter measures oxygen in blood mentions as an older method, uses exactly the same logic with a dye instead of cold.
If you want more method-of-mixtures and heat-flow problems like this, the Fermi Problem Library has a step-by-step solver that explains the reasoning at each step.
Quick checklist
Thermodilution: inject a known amount of cold, measure how much the flowing blood warms it.
Heat gained by saline = heat lost by blood, summed over the dip.
Q = V_i (T_b − T_i) K / ∫ΔT dt; K ≈ 1.08 with approximate data.
Larger flow gives a smaller, shorter dip and a smaller area.
10 mL at 0 °C and 5 L/min gives about 4.79 K s; room-temperature saline gives about 40% of that.
The method assumes full mixing, one pass, no stray heat loss and steady flow.