An optical coherence tomography (OCT) scanner can show the layers at the back of your eye, each a few micrometres apart, using nothing but light and a mirror. It works because of a wave-optics idea that most textbooks mention only in passing: light with a wide spread of wavelengths interferes only over a very short path difference. OCT turns that short window into a ruler. This post works out how short the window is, why a broad light source makes it shorter, and why a single-colour laser would be useless for picking out one depth.
This is a physics explainer, not medical advice. An OCT scan is read and interpreted by an eye doctor.
Why does broadband light interfere only over a short distance?
Two waves give steady fringes only if their phase difference stays fixed. Our post on why two independent light sources cannot produce interference explains the phase jumps that spoil this between separate sources. OCT is about something else: a single source, split in two, with one half delayed.
Think of broadband light as many wavelengths at once. At zero path difference, every wavelength has a bright fringe in the same place. Move away from zero and each wavelength's fringes drift by a different amount. After a short extra path, the bright fringes of some colours sit on the dark fringes of others, and the pattern washes out to an even grey. This is exactly why white-light fringes in YDSE turn coloured and fade after a few orders.
The wider the spread of wavelengths Δλ, the sooner the colours fall out of step. The path difference over which fringes survive is called the coherence length.
What is the coherence length formula?
For a source with centre wavelength λ and a bandwidth Δλ (full width at half maximum), the fringe envelope against path difference has a width of about
l_c ≈ 0.88 λ²/Δλ (path difference).
The exact number in front depends on the shape of the spectrum; 0.88 is for a smooth bell-shaped (Gaussian) one. What matters for exams is the scaling: l_c ∝ λ²/Δλ. Broad bandwidth means short coherence; narrow bandwidth means long coherence.
In OCT the light goes into the eye and back, so moving a reflecting layer by a depth d changes the path by 2d. The depth window is therefore half the path window:
Δz ≈ 0.44 λ²/Δλ (axial resolution, in air).
This is the standard formula for OCT depth resolution, stated with the 0.44 factor in a peer-reviewed paper by Kolb and colleagues in Biomedical Optics Express (2018). The same paper notes that axial (depth) and transverse (sideways) resolution are decoupled: depth resolution comes from the light's bandwidth, while sideways resolution comes from how tightly the optics focus the beam.
How does an OCT scanner use this to measure depth?
The layout is a Michelson interferometer, the same arrangement described in our post on Rainer Weiss and the LIGO interferometer, but with broadband light instead of a laser.
A beam splitter sends half the light to a reference mirror and half into the eye.
Each layer of the retina reflects a little light back.
The two returning beams meet at a detector.
Fringes appear only for light from the layer whose path matches the reference path to within Δz. Light from every other depth adds only a steady background.
Moving the reference mirror moves the matching depth. Recording fringe strength against mirror position gives a depth profile; sweeping the beam sideways builds a cross-section image.
That is the simplest version, called time-domain OCT. Many modern instruments keep the mirror still and read all depths at once from the spectrum of the returning light (spectral-domain OCT), or sweep the colour of a tunable laser (swept-source OCT). The depth resolution is still set by the total bandwidth covered.
How fine a depth can an OCT scan resolve?
Take a source centred at 840 nm, a wavelength used in retinal spectral-domain OCT, with a round, illustrative bandwidth of 50 nm.
Write the formula. Δz = 0.44 λ²/Δλ.
Square the wavelength. λ² = (840 × 10⁻⁹ m)² = 7.06 × 10⁻¹³ m².
Divide by the bandwidth. 7.06 × 10⁻¹³ ÷ 50 × 10⁻⁹ = 1.41 × 10⁻⁵ m.
Multiply by 0.44. Δz = 6.2 × 10⁻⁶ m = 6.2 µm in air.
Correct for the eye. Light travels slower in tissue, so the same optical delay is a shorter real distance. Divide by the refractive index; with n ≈ 1.34, Δz ≈ 4.6 µm in tissue.
Check units. m²/m = m, so the answer is a length.
That is close to the 5 to 7 µm axial resolution of standard commercial instruments quoted by Lee and colleagues in Translational Vision Science & Technology (2020), who also used n = 1.3375 for the eye. The same paper describes an 840 nm research system with a 150 nm bandwidth reaching 2.7 µm in tissue: three times the bandwidth, a much finer depth.
For a sense of scale, take a retina about 250 µm thick (a round, illustrative value). With 4.6 µm depth steps, the scan can separate about 54 slices through it.
Why can't a single-colour laser pick out one depth?
Run the same formula the other way. A helium–neon laser at 632.8 nm with a bandwidth of about 1 pm (an illustrative figure; real lasers vary widely) has a depth window of 0.44 × (632.8 nm)² ÷ 1 pm ≈ 0.18 m. Light from every layer of the retina, and from the cornea and lens too, would interfere at once. You could not tell which layer a fringe came from.
That long coherence is what makes a laser the right choice for LIGO, which needs interference over kilometres of arm length, and the wrong choice for a fixed-wavelength depth scanner. Swept-source OCT does use a laser, but its wavelength is swept across a wide band, so the bandwidth it covers is still what sets depth resolution.
The same formula also explains the classroom fact. White light centred near 550 nm with Δλ ≈ 300 nm gives l_c ≈ 0.88 × (550 nm)² ÷ 300 nm ≈ 0.9 µm of path difference, under two wavelengths. That is why white-light fringes are visible only near the central one.
The Concept Library has the interference and coherence ideas behind this, and the Simulation Lab has interactive simulations of fringes you can adjust.
Quick checklist
Fringes survive over a path difference l_c ≈ 0.88 λ²/Δλ; OCT depth resolution is half that, Δz ≈ 0.44 λ²/Δλ.
Broad bandwidth gives fine depth resolution; narrow bandwidth gives long coherence.
Divide by the refractive index for depths inside tissue.
840 nm with 50 nm bandwidth gives 6.2 µm in air, about 4.6 µm in the eye.
Depth resolution comes from bandwidth; sideways resolution comes from focusing.
An OCT image is a measurement; an eye doctor interprets it.