A water pump on a 230 V supply draws 5 A. Multiply the two and you get 1,150. Put a wattmeter in the same circuit and it reads 920 W. Neither meter is broken. A wattmeter's pointer settles at the average of the instantaneous power, and for a motor that average is smaller than V × I. Here is how two coils and a spring do that, with the numbers for the pump.
What is inside an electrodynamometer wattmeter?
The classic laboratory wattmeter is the electrodynamometer type. It has two sets of coils.
The current coil is fixed. It is made of a few turns of thick wire and is connected in series with the load, so the full load current i flows through it.
The pressure coil (also called the voltage coil) is light and free to turn on a spindle. It is connected across the supply through a large series resistance, so the small current through it is proportional to the voltage v.
A spring provides the controlling torque, and a damper stops the pointer swinging.
The INFLIBNET course chapter on the measurement of power and wattmeters gives this construction, with a pressure-coil current of roughly 10 to 50 mA.
There is no iron in the classic design, so it works on DC too, where it simply reads V × I.
Why is the torque proportional to power?
The fixed coil makes a magnetic field proportional to the load current i. The moving coil carries a current proportional to v. The force on a current-carrying wire in a field is proportional to the current times the field, so the torque on the moving coil goes as the product of the two currents. This is the same physics as the force between two parallel currents: two currents push or pull on each other with a force proportional to i₁i₂.
So at every instant, torque ∝ i × (v/R) ∝ v × i = p, the instantaneous power. When both currents reverse together, the torque does not reverse, because the product of two negatives is positive. That is why the meter works on AC at all.
Why does the pointer show average power, VI cos φ?
On 50 Hz mains, p = vi swings up and down at 100 Hz, twice the supply frequency. The pointer and coil have far too much inertia to follow a torque that changes a hundred times a second. They settle where the spring torque balances the average torque, so the deflection is proportional to the average power. The same chapter of the INFLIBNET measurements course states this: the moving system cannot follow the rapid changes in torque.
For sinusoidal v and i with rms values V and I and a phase difference φ, that average is P = VI cos φ. Why the cos φ appears, and what power factor means, is set out in power factor in AC circuits, so it is not repeated here. The point here is that the wattmeter does the averaging mechanically. You never have to know φ.
An ordinary AC voltmeter or ammeter reports a size and throws away the timing. How a mechanism decides what a meter reports is the theme of why a cheap multimeter misreads a dimmer. A wattmeter is the meter whose mechanism multiplies before it averages.
Worked example: why does the pump read 920 W?
The pump runs on 230 V rms and draws 5 A rms. Its current lags the voltage by φ = 36.87°, so cos φ = 0.8.
Apparent power. V × I = 230 × 5 = 1,150 VA. This is what a voltmeter and an ammeter together suggest.
Wattmeter reading. P = VI cos φ = 1,150 × 0.8 = 920 W.
Instantaneous power. For sine waves, p = VI cos φ − VI cos(2ωt − φ). It swings by ±VI = ±1,150 W about the average of 920 W.
Its range. Maximum 920 + 1,150 = 2,070 W. Minimum 920 − 1,150 = −230 W.
When is p negative? Whenever v and i have opposite signs. In each half cycle that lasts for φ out of 180°, so the fraction is 36.87/180 = 20.5% of every cycle. During those moments the motor's magnetic field hands energy back to the supply.
Now change the load and keep V × I = 1,150 VA:
φ = 60°: P = 1,150 × 0.5 = 575 W.
φ = 85° (almost a pure choke): P = 1,150 × 0.0872 = 100 W.
The ammeter and voltmeter readings have not changed, yet the wattmeter has dropped from 920 W to 100 W. That is why the instrument exists.
Why do wattmeters struggle at low power factor?
Two things go wrong when φ is close to 90°.
First, the reading is small. At φ = 85° the coils carry full current, yet the pointer shows only 100 W.
Second, the pressure coil is not a perfect resistor. It has a small inductance, so its current lags the supply voltage by a small angle β. The INFLIBNET chapter on low power factor wattmeters gives the resulting error as about VI sin φ tan β. It makes the meter read high when the current lags the voltage and low when the current is ahead of it.
To see the size, take an illustrative tan β = 0.001 (a round value, not a figure for a real instrument):
At φ = 36.87°: error = 1,150 × 0.6 × 0.001 = 0.69 W, which is 0.075% of 920 W.
At φ = 85°: error = 1,150 × 0.996 × 0.001 = 1.15 W, which is 1.1% of 100 W.
The same small phase error is fifteen times worse in percentage terms at low power factor. That is why special low power factor wattmeters exist. The same source says they use a weaker control spring so that full-scale deflection is reached at power factors as low as 0.1, and a capacitor across part of the series resistance to cancel the coil's inductance.
Are wattmeters still built this way?
Mostly in teaching labs. Modern wattmeters and home electricity meters are electronic. They sample v and i many times in every cycle, multiply each pair and average the products. The method is the same: multiply first, then average.
A note on safety: measurements on 230 V mains are for trained people with the right equipment. In a school lab, the same ideas are shown safely with a low-voltage AC supply. If you want to see how the phase angle changes the shape of p = vi, try it in the Simulation Lab, and use the Fermi Problem Library for AC power problems with a step-by-step solver that explains the reasoning.
Quick checklist
Current coil: fixed, in series, carries the load current.
Pressure coil: moving, across the supply through a large resistance, carries a current proportional to v.
Torque ∝ product of the two currents ∝ instantaneous power vi.
Inertia averages the 100 Hz swings, so the reading is VI cos φ.
230 V, 5 A, cos φ = 0.8: reads 920 W; p swings from −230 W to 2,070 W.
Power factor from readings = wattmeter reading / (V × I).
Low power factor makes readings small and phase errors large.