An X-ray beam aimed at a tumour 15 cm deep has to pass through the healthy tissue in front of it, and then it carries on through the tissue behind it. A proton beam behaves differently. It deposits only a modest dose on the way in, gives most of its energy in the last centimetre or so, and then stops. That sharp rise at the end is the Bragg peak, and it is the whole physical reason proton therapy exists. Here is where the peak lands, why, and why relativity matters.
This is a physics explainer, not medical advice. Whether proton or X-ray treatment suits a particular patient is decided by oncologists and medical physicists, case by case.
How does an X-ray beam deposit its dose?
High-energy X-ray photons are absorbed in single events, mostly by Compton scattering. The beam thins out rather than slowing down. Once the dose has built up in the first centimetre or two under the skin, it falls slowly and roughly exponentially with depth, and the beam leaves the body on the far side. The basic idea of treating with X-rays in fractions is covered in how radiotherapy works and how it differs from chemotherapy, so we will not repeat it here.
A charged particle is different. A proton loses its energy a little at a time, in a very large number of collisions with atomic electrons. So it has a definite range: after a certain depth it has nothing left to lose.
Why does a proton dump most of its energy at the end?
The energy a fast charged particle loses per centimetre (the stopping power) rises roughly as 1/v². A slow proton spends longer near each electron it passes, so it gives each one a bigger push. As the proton slows, it loses energy faster, which slows it further. The process runs away in the final millimetres.
The NIST PSTAR database gives the stopping power of liquid water for protons:
At 150 MeV: about 5.4 MeV per g/cm², which is 5.4 MeV per cm of water.
At 10 MeV: about 45.6 MeV per cm, over 8 times more.
At 1 MeV: about 261 MeV per cm, 48 times more.
PSTAR also gives the range of a 10 MeV proton in water as 0.12 g/cm², about 1.2 mm. So a proton that entered at 150 MeV gives up its last 10 MeV, one fifteenth of its energy, in the final 1.2 mm of its path. Add up millions of protons for a low, flat entrance dose followed by a tall, narrow peak.
How deep does a proton beam go?
A useful fit, called the Bragg–Kleeman rule, gives the range R in water from the kinetic energy E:
R ≈ 0.0022 × E¹·⁷⁷ (R in cm, E in MeV)
It is an approximation. For precise values, use the PSTAR table, which quotes the CSDA range (the path length if the proton slowed down smoothly) in g/cm². For water, with density 1 g/cm³, that number is also the depth in cm.
Worked example: range and energy for a tumour
Range at 70 MeV. 0.0022 × 70¹·⁷⁷ = 4.1 cm. PSTAR: 4.08 g/cm².
Range at 150 MeV. 0.0022 × 150¹·⁷⁷ = 15.6 cm. PSTAR: 15.77 g/cm², about 15.8 cm.
Range at 230 MeV. 0.0022 × 230¹·⁷⁷ = 33.3 cm. PSTAR: 32.95 g/cm², about 32.9 cm.
Invert the rule to find the energy for a given depth: E = (R/0.0022)^(1/1.77).
Tumour 10 cm deep. E = (10/0.0022)^0.565 = 117 MeV. PSTAR gives 10.2 g/cm² at 117 MeV, so the fit is within 2%.
Tumour 20 cm deep. E = (20/0.0022)^0.565 = 172 MeV. PSTAR gives 20.0 g/cm² at 172 MeV.
Because the exponent 1.77 exceeds 1, doubling the depth needs much less than double the energy. Real tissue is not pure water, so clinical planning converts each tissue to a water-equivalent depth using scans.
Why do therapy protons need relativity?
A proton's rest energy is mc² = 938.3 MeV. At 150 MeV the kinetic energy is 16% of that, which is too large to treat classically.
Lorentz factor. γ = 1 + T/mc² = 1 + 150/938.3 = 1.160.
Speed. v/c = √(1 − 1/γ²) = √(1 − 1/1.160²) = 0.507. The proton moves at about half the speed of light.
Classical answer. v/c = √(2T/mc²) = √(2 × 150/938.3) = 0.566.
Error. 0.566/0.507 = 1.12, so the classical formula is about 12% too high.
At 230 MeV. γ = 1.245 and v = 0.596c, while the classical formula gives 0.700c, about 18% too high.
At 70 MeV the classical error is still about 5%. If you have practised proton and alpha speeds through the same voltage, that classical method is fine for kilovolts. At therapy energies, switch to γ.
What machine accelerates the protons?
A cyclotron is the usual choice. The simple cyclotron in maximum kinetic energy in a cyclotron assumes the orbit frequency f = qB/2πm stays fixed. Relativistically the frequency is qB/2πγm, so at 230 MeV it has fallen by a factor of 1.245, about 20%. A fixed-frequency, fixed-field machine would fall out of step with the protons.
Therapy machines solve this in two ways. An isochronous cyclotron makes the magnetic field grow with radius so that γm/B, and so the orbit time, stays constant. A synchrocyclotron lowers the RF frequency as each bunch of protons speeds up. Some centres use synchrotrons instead. A review of cyclotrons for cancer therapy by Yves Jongen of IBA describes all three, and notes that Robert Wilson first pointed out the use of the Bragg peak for radiotherapy in a 1946 paper.
To cover a tumour that is several centimetres thick, the beam energy is stepped through a set of values. Each energy puts its peak at a different depth, and together they build a flat high-dose region called a spread-out Bragg peak. For ideas that are easier to see than read, the Simulation Lab has interactive simulations.
Does the dose stop completely after the peak?
Almost, but not exactly. Three effects blur the edge.
Range straggling. Energy loss is a random process, so identical protons stop at slightly different depths, spread by roughly 1% of the range.
Nuclear reactions. A small fraction of protons, around 1% per cm of water, hit nuclei instead of electrons. These produce secondary particles, including neutrons, that carry some dose beyond the peak.
Uncertainty in the range itself. Converting tissue to water-equivalent depth is not perfect, so planners allow margins.
So the dose beyond the peak is small, but it is not zero. The practical advantage over X-rays is the much lower exit dose, not a perfect cut-off.
Quick checklist
X-rays thin out exponentially; protons slow down and stop.
Stopping power rises roughly as 1/v², which produces the Bragg peak.
Range in water: about 4.1 cm at 70 MeV, 15.8 cm at 150 MeV, 32.9 cm at 230 MeV (PSTAR, g/cm²).
Bragg–Kleeman fit: R ≈ 0.0022 E¹·⁷⁷; invert it to find the energy for a given depth.
Above about 10 MeV, use γ = 1 + T/mc² for proton speed.
Therapy accelerators are isochronous cyclotrons, synchrocyclotrons or synchrotrons.
Treatment decisions belong to oncologists; this is the physics only.