On this page
Your vernier caliper has a least count of 0.02 mm and your screw gauge 0.01 mm. You might think that is the whole story: one is twice as fine as the other. But there is a second, bigger difference in how they are built. In a screw gauge the scale lies on the same line as the object. In a caliper it does not. That one design choice, named after the German physicist Ernst Abbe, decides how much a tiny tilt can spoil your reading.
What is Abbe's principle?
The measuring-instrument maker Mitutoyo's Quick Guide to Measurement states Abbe's principle as: "maximum accuracy is obtained when the scale and the measurement axes are common". In plain words, put the scale on the same straight line as the length you are measuring.
If the two lines are separated by a distance d, called the Abbe offset, any small rotation θ of the moving part shifts the measuring point by d × tanθ. The scale does not see this shift, because the scale is not where the shift happens. That unseen shift is the Abbe error:
Abbe error ε = d × tanθ ≈ d × θ (θ in radians)
It is a first-order error: it grows in direct proportion to θ. The same guide notes that play in a guide, straightness errors or a change in measuring force can all make θ vary.
Why does a vernier caliper break Abbe's principle?
Look at a caliper. The main scale and the vernier run along the beam at the top. The object sits between the jaws, below the beam. The distance from the scale line down to where the jaws touch the object is the Abbe offset d. If you hold a small object deep in the jaws, d can be several centimetres.
The sliding jaw rides on the beam with a little clearance. If you squeeze the thumb roller too hard, or the guide is worn, the slider can rock by a tiny angle. The jaw tip then moves by d × tanθ, while the vernier, sitting on the beam, barely moves. Mitutoyo's guide lists this as moving jaw tilt error and says it follows the same formula as errors caused by not conforming to Abbe's principle.
How to take a caliper reading and correct for zero error is covered in reading a vernier calliper and screw gauge, with zero error. Abbe error is a different systematic error from zero error, whose sign trips up many students: it does not show up when you close the jaws, because the tilt depends on how you hold and squeeze.
Why does a screw gauge barely notice a tilt?
In a screw gauge the anvil, the object, the spindle and the screw all lie on one axis. The scale reading comes from the screw itself, along that same axis. So the Abbe offset is essentially zero, and the first-order term d × θ vanishes.
What is left is a second-order effect called cosine error. If the length you want, L, lies at a small angle θ to the measuring axis, you measure L cosθ instead of L. The error is
cosine error = L(1 − cosθ) ≈ Lθ²/2
Because θ is squared, a small angle gives a very small error. Mitutoyo's guide shows the same cosine effect for dial test indicators, where the true value equals the indicated value × cosθ, and advises keeping the angle between the movement directions as small as possible.
Worked example: the same 1 mrad tilt in both instruments
A tilt of 1 milliradian (mrad) is 0.001 rad, or about 0.057°. That is the angle a 50 mm jaw makes if its tip is 0.05 mm out of line. The values below are illustrative.
Caliper, object held 40 mm below the scale line.
Abbe offset d = 40 mm, θ = 0.001 rad.
ε = d × tanθ = 40 mm × 0.001 = 0.040 mm = 40 µm.
Compare with the least count: 40 µm ÷ 20 µm = 2. The reading is off by two whole least counts.
With θ = 0.5 mrad, ε = 20 µm, one least count.
Hold the same object near the beam, d = 10 mm, and the 1 mrad error drops to 10 µm.
Mitutoyo's guide gives its own example: a jaw slope of 0.01 mm in 50 mm (0.2 mrad) with jaws 40 mm deep gives an error of 40 × 0.01 ÷ 50 = 0.008 mm at the jaw tip.
Screw gauge, 25 mm object, same 1 mrad misalignment.
L = 25 mm, θ = 0.001 rad.
Cosine error = Lθ²/2 = 25 mm × (0.001)²/2 = 1.25 × 10⁻⁵ mm = 0.0125 µm.
That is about one-thousandth of a 10 µm least count, far too small to see.
Even at 1°, a large misalignment, the error is 25 × (1 − cos 1°) = 0.0038 mm = 3.8 µm, still below one least count.
Check units. Offset in mm times an angle in radians (a pure number) gives mm. In the cosine error, θ² is also a pure number, so the answer is in mm.
Does this mean calipers are inaccurate?
No. A caliper used as intended stays within its stated tolerance. The same Mitutoyo guide lists the maximum permissible error under the Japanese standard JIS B 7507:2016: for a caliper with 0.02 mm graduation, ±0.02 mm for lengths up to 50 mm. That allowance already covers the small tilts of a caliper in good condition used with light, steady force. Our 40 µm example would exceed it, which is a sign of a worn guide or a heavy thumb, not of normal use.
What Abbe's principle tells you is how to use a caliper well:
Measure with the object as close to the beam as the jaws allow, to keep d small.
Use light, even pressure on the thumb roller so the slider does not rock.
Read the scale straight on, so you do not add parallax; see the correct position of the eye while reading a scale.
When you need better than about 0.02 mm, use a screw gauge.
For practice with least count and systematic error questions, the Concept Library covers every concept in physics, including units and measurement.
Quick checklist
Abbe's principle: scale and measuring axis should be on one line.
Abbe error = offset × tanθ, first order in θ.
A vernier caliper has an offset (scale on the beam, object in the jaws). A screw gauge does not.
Cosine error = L(1 − cosθ) ≈ Lθ²/2, second order in θ.
1 mrad tilt: about 40 µm on a caliper at 40 mm offset, about 0.01 µm on a 25 mm screw gauge reading.
Keep the object near the beam and squeeze lightly.